Onboarding
July 14, 2026
2 min read

How accurate is Shen AI?

How accurate is Shen AI?

Reliable health measurements start with reliable validation. Shen AI is continuously evaluated against medical reference devices to understand how closely its measurements match established clinical methods. These validation studies help us improve the SDK, measure progress over time, and give customers a transparent view of its performance.

This article summarizes the results of our latest clinical validation study. For the complete methodology, statistical analysis, and detailed performance tables, see the Clinical validation report.

How do we validate Shen AI?

Our latest clinical validation study included 132 adult participants measured over two clinical visits. During each measurement, participants were recorded with a standard smartphone camera while medical reference devices measured the same vital signs at the same time. This allowed researchers to compare every Shen AI result with a trusted clinical reference.

The study evaluated four key measurements:

- Heart Rate
- Heart Rate Variability (HRV)
- Breathing Rate
- Blood Pressure

Heart Rate

Heart Rate produced the strongest results in the study. Using a standard 60-second measurement, Shen AI achieved an average error of just 0.14 beats per minute with an almost perfect correlation (0.99) compared to ECG. Even after shortening the measurement to 30 seconds, performance remained very similar.

That's one of the reasons we were able to reduce the standard measurement time from 60 seconds to 30 seconds without compromising the user experience.

Duration N MAE (bpm) MAPE (%) RMSE (bpm) SD (bpm) Mean bias (bpm) Correlation
60 s 784 0.14 0.20 0.41 0.41 −0.05 0.99
45 s 759 0.16 0.24 0.45 0.45 −0.04 0.99
30 s 780 0.31 0.46 0.89 0.89 −0.07 0.99

Table 1: Summary of heart rate accuracy (Shen AI vs ECG)

Heart Rate Variability

Heart Rate Variability is much more difficult to measure than Heart Rate because it looks at tiny differences between individual heartbeats rather than simply counting them.

Even so, Shen AI achieved a correlation of 0.95 with ECG measurements in the clinical study. Because HRV requires more data to produce stable results, it's always calculated using the full measurement window.

Duration N MAE (ms) MAPE (%) RMSE (ms) SD (ms) Mean bias (ms) Correlation
60 s 776 4.02 18.36 6.66 6.54 −1.29 0.95

Table 2: Heart rate variability (SDNN) accuracy

Breathing Rate

Breathing is naturally slower than the heartbeat, so the SDK needs to observe more breathing cycles before it can produce a reliable estimate.

For that reason, breathing rate benefits from longer recordings. In our study, the average error for a 60-second measurement was 1.4 breaths per minute, while shorter measurements still produced results useful for many wellness applications.

Duration N MAE (bpm) MAPE (%) RMSE (bpm) SD (bpm) Mean bias (bpm) Correlation
60 s 713 1.40 8.50 2.31 2.31 −0.22 0.77
45 s 687 1.32 8.67 2.22 2.22 0.06 0.77
30 s 708 1.44 9.64 2.45 2.45 0.08 0.74

Table 3: Breathing rate accuracy

Blood Pressure

Our clinical study evaluated both calibrated and non-calibrated blood pressure estimation. The results showed that individual calibration can improve agreement with traditional cuff measurements, and we continue to refine these algorithms through ongoing research and clinical validation.

Parameter N MAE MAPE RMSE SD Bias r
SBP (90–170) 1184 10.18 9.41 12.62 11.76 4.60 0.38
DBP (60–100) 1071 6.22 8.96 7.71 7.27 2.57 0.34

Table 4: Blood pressure accuracy

Note: MAE, RMSE, SD, and Bias are in mmHg; MAPE in %; r = Pearson correlation.

In summary

Clinical validation is an ongoing part of how Shen AI is developed. Every study helps us better understand the strengths of the technology, improve existing algorithms, and validate new capabilities before they reach production.

The results presented here demonstrate strong agreement with established medical reference devices across the vital signs measured by Shen AI. For customers evaluating the SDK, they provide an independent benchmark of how the technology performs under controlled clinical conditions. If you want to test Shen AI’s accuracy yourself, follow the recommendations listed in this guide.

Share this post

More blog posts

Sagittis et eu at elementum, quis in. Proin praesent volutpat egestas sociis sit lorem nunc nunc sit.

Blog
June 3, 2026
4 min read

Beyond annual checkups: A different approach to chronic care screening

Read more
Blog
May 13, 2026
5 min read

What’s changing in women’s health - and what still isn’t

Despite growing investment in femtech, women’s health is still shaped by data gaps and system-level bias. This discussion explores how better data, thoughtful AI, and more intentional product design can move from tracking to real understanding.

Read more
Blog
May 1, 2026
8 min read

The FDA's 2026 general wellness update and what it means for health monitoring

The FDA’s 2026 update expands what qualifies as general wellness, but applying it correctly is where most teams struggle. This webinar recap breaks down the key changes, common compliance pitfalls, and how product, marketing, and AI decisions shape regulatory risk in digital health.

Read more